021 8611 259
SALES  load cell yahoo messenger
SALES  load cell yahoo messenger
Home   »   Produk   »   Strain Amplifier
  • weigh bridge load cell
  • Banner RLC1 kecil
  • Banner s-beam load cell
  • Banner RLC2 kecil
  • Commpresion Load Cell
  • Banner RLC3 kecil
  • Load Cell application
  • Banner RLC4 kecil
  • Torque sensor
  • Banner 4
  • Banner RLC1
  • Banner RLC2
  • Banner RLC3
  • Banner RLC4
  • Banner Force Gauge
  • Banner Cap Torque Tester
  • Banner Torque Tester

Static Strain Digital Indicator PSD-704

 
 
 

Untuk memesan, silahkan tekan tombol ORDER. Anda bisa memesan barang secara Online.

 

Anda juga bisa memesan via telephone. Silakan hubungi sales consultant, Hotline kami di, 021 8611 259 sales Consultant kami akan menjelaskan detail produk dan harga terbaru. atau email ke alamat sales@rajaloadcell.com

FEATURES

  Indicator for strain gage or strain gage applied transducer
  Check strain gage based transducers in mV/V.
  Measure I/O resistance of strain gage based transducers. Insulation resistance between main body and each cable wire is also available.
  Applied RS-232C interface as a standard.

 

SPECIFICATIONS

Bridge power supply DC 2V±0.02 V Within 35 mA
Target for measurement Strain measurement, Measurement on output of strain gage applied transducer, Meadurement on resistance value.
Measuring method Deflection method
(under measuring on strain or output of strain gage application)
Temperature efficient Zero point ±0.2x10-6 in strain/℃, Sensitivity ±0.01 %F.S./℃
(After 15 min. of warming up time with the range of x1)
Efficient due to time Zero point ±0.2x10-6 in strain/8 h, Sensitivity ±0.01 %F.S./8 h
(Temperature variation within ±2 oC with the range of x1)
A/D sampling 4 times/s
Display Dot matrix type liquid crystal display (with back light LED type)
Specification for strain measurement  
Measuring method 1 gage 2 wire method
Strain Gage : 120 Ω, 240 Ω, 350 Ω
1 gage 3 wire method
Strain Gage : 120 Ω, 240 Ω, 350 Ω
2 gage method
Strain Gage : 60 Ω to 1 000Ω
4 gage method (Strain gage applied transducer)
Strain Gage : 60 Ω to 1 000Ω
Measuring range x1 range : ±40 000x10-6 in strain
4GH range : ±3.500 0 mV/V
(±7 000.0x10-6 in strain)
(4 gage method, Only for strain gage applied transducer)
Resolution x1range : 1 x10-6 in strain
4GHrange : 0.000 1 mV/V(0.1x10-6 in strain)
(4 gage method, Only for strain gage applied transducer)
Accuracy x1range : ±0.08 %F.S.±1 digit
4GHrange : ±0.08 %F.S.±2 digit
(4 gage method, Only for strain gage applied transducer)
Operating temp./humidity range Temperature 0 ℃ to 50 ℃, Humidity 85 %RHor less(Non condensing)
Power supply voltage AC90 V to 132 V(in the usage of AC adapter) 50/60 Hz
or battery drive by Ni-Cd battery.
Power consumption Approx. 0.8 VA(When AC adapter is used.)
Outline dimensions(WxHxD) 210 mmx148.5 mmx40 mm(excludes protruding parts)
Weight Approx. 1.7 kg