DC type dynamic strain amplifier DAS-406B
Untuk memesan, silahkan tekan tombol ORDER. Anda bisa memesan barang secara Online.
Anda juga bisa memesan via telephone. Silakan hubungi sales consultant, Hotline kami di, sales Consultant kami akan menjelaskan detail produk dan harga terbaru. atau email ke alamat sales@rajaloadcell.com
FEATURES
- DC ~ 200 kHz Characteristics in the frequency
- Standard equivalent strain amplifier with four digit digital switch setting
- 3 1/2 digit digital monitor
SPECIFICATIONS
| Measuring point | 1 point/1 unit |
|---|---|
| Applicable gage resistance | 60 Ω ~ 1kΩ |
| Gage factor | 2.00 |
| Bridge power supply | DC2 V(DC1, 4, 10 V Changeable to) with remote sensing function(50 mA or less) |
| Balancing adjustable range | Resistance section Approx. ±1 % (Gage resistance 120 Ω Approx. ±5 000x10-6 in strain) |
| Balancing adjustable method | Auto balance, Auto balance time Approx.1 s, Accuracy ±0.1 %F.S. |
| Sensitivity | 0.5 V output at the input of 100 x 10-6 in strain (When bridge power supply is DC2 V) |
| Output | MONITOR ±10 V, ±5 mA (Load resistance2 kΩor more) OUTPUT ±10 V, ±30 mA (Load resistance 330 Ωor more) Output resistance10 Ω or less Output capacitive load0.1 μF or less |
| Non-linearity | ±0.01 % F.S. |
| Standard equivalent strain(CAL) | Set with digital switch ±1 ~ ±9 999x10-6 in strain , Accuracy ±0.2 %F.S. or less at each point |
| Sensitivity adjustable range(ATT) | 1, 1/2, 1/5, 1/10, 1/20, 1/50, 1/100, 0 Accuracy ±0.2 %F.S. at each point GAIN volumu valiable range 1 to 1/2.5 times continuous vary(VAR) |
| Frequency response range | DC to 200 kHz +1 dB, -3 dB (Filter at W/B) |
| Low-pass filter | 10 Hz, 100 Hz, 1 kHz, 10 kHz, W/B Third bessel type |
| S/N ratio | 46 dBp-p(ATT=1) with W/B condition 55 dBp-p with 10 Hz to 10 kHz |
| Display | Output voltage display 0.00 to ±10.00 Digital display (Green LED) Sampling Approx.4 times/s 0.00 or -0.00 flashes when output voltage is in non-linearity zone |
| Temperature efficient | Zero shift ±0.1x10-6 in strain / ℃ Sensitivity variation ±0.01 %F.S./℃ |
| Efficient due to time | Zero shift ±5x10-6 in strain / 8 h Sensitivity variation ±0.05 %F.S./8 h |
| Efficient due to power supply | Zero shift ±1x10-6 in strain / variation of power supply voltage within ±10 % Sensitivity valliation ±0.01 %F.S. /variation of power supply voltage within±10 % |
| Operating temp./humidity range | Temperature :-10 ℃ to 50 ℃, Humidity 85 %RHor less (Non condensing) |
| Power supply voltage | AC100 V±10 V, DC12 V 50/60 Hz |
| Power consumption | Approx. 10 VA(at AC100 V), Options DC12 V Approx. 0.8 A |
| Outline dimensions(WxHxD) | 42 mmx176 mmx284.4 mm(excludes protruding parts) |
| Weight | Approx. 1.5 kg |









.jpg)